Rules · QAQC
R100 · CRM outside the fail limit
Warning QAQC
A certified reference material has a known grade. When the lab reports it outside the outer limit (3 SD, or ±10% without an SD), the batch it was in cannot be trusted for that method: the verdict is fail, the certificate-method is held, and a re-assay set is needed.
What it checks
For each CRM on a certificate, compares each result with the library's certified value for the same element and method, both in ppm. A result further from it than the outer limit is reported.
Why it matters
A CRM has a known grade. When the lab misses it by that much, the batch cannot be trusted for that method: the verdict is fail, the certificate and method are held (R97), and a re-assay set is needed.
What to do
First check that the right CRM code is on the sample, since swapped standards are common. If the code is right, ask the lab to re-assay the batch around the failed standard.
Settings
| Setting | Default | What it means |
|---|---|---|
| Use SD limits when the library has an SD (else ± % of certified) | On | When on (default), limits are set in standard deviations wherever the CRM library has an SD for the value; otherwise they are a percentage of the certified value. |
| Outer (fail) limit in SD | 3 SD | The fail limit, in standard deviations from the certified value (default 3). |
| Outer (fail) limit in % of certified | 10 % | The fail limit as a percentage of the certified value, used when there is no SD or SD limits are off (default 10%). |
| Inner (warning) limit in SD | 2 SD | The warning limit, in standard deviations from the certified value (default 2). This rule does not use it; R101 does. |
| Inner (warning) limit in % of certified | 5 % | The warning limit as a percentage of the certified value, used when there is no SD or SD limits are off (default 5%). This rule does not use it; R101 does. |
From Collar 1.1.0's rule catalogue. The same text is in the app under Help.